Error lists/Troubleshooting 3-11
Centrifuge 5804/5804 R and 5810/5810 R
3 Error lists / Troubleshooting
Error 36 EEPROM not yet initialized. Reinitialize EEPROM with service
program 5.
Change PCB "Control".
Error 37 Voltage version of frequency converter
and Centrifuge do not correspond.
undefined
display, no
background
illumination
Display module or 26-pole cable
defective, 5 V power supply faulty
Check all cables.
Change PCB Control pad or PCB
Control.
Display
Cause
MSTM/0408