Error lists/Troubleshooting 3-11
Centrifuge 5804/5804 R and 5810/5810 R
3 Error lists / Troubleshooting
Error 36 EEPROM not yet initialized. – Reinitialize EEPROM with service
program 5.
– Change PCB "Control".
Error 37 Voltage version of frequency converter
and Centrifuge do not correspond.
undefined
display, no
background
illumination
Display module or 26-pole cable
defective, 5 V power supply faulty
– Check all cables.
– Change PCB Control pad or PCB
Control.
Display
Cause
MSTM/0408