Centrifuge 5804/5804 R and 5810/5810 R
3 Error lists / Troubleshooting
3-12 Error lists/Troubleshooting
Warnings / Error displays
(are not stored in the EEPROM)
Display Cause
Voltage
false
W-Error 27
Mains voltage of FC incorrect.
Check voltage version.
SPEED W-Error 29
‘SPEED’ – the set rotational speed is faster
than the nominal rotational speed of the rotor
in use.
The device stops the run after rotor
recognition.
Enter new nominal speed.
200 rpm New rotor was used in devices with older soft-
ware
Install new software and carry out
imbalance adjustment.
IMBAL W-Error 30
‘IMBAL’ – Imbalance at < 2000 1/min.
The device brakes.
No imbalance calibration was carried out after
installing rotor A-4-81.
Check rotor and loading. Test buckets
and adapters with a scale.
Clean pegs and nuts of the rotor/
bucket and grease slightly with
Eppendorf pivot grease.
Test another rotor.
Carry out imbalance calibration.
Rotor See Error 15.
Caution with open lid – locking mechanism is actuated! Risk of squeezing!