Centrifuge 5804/5804 R and 5810/5810 R
3 Error lists / Troubleshooting
3-26 Error lists/Troubleshooting
Is the refrigerating system tight? Does the valve switch? Repair if necessary and refill cooling agent
Quantity – cooling agent R134A
5804 R 400 g
5810 R 470 g
Repairing the refrigerating system
The spare parts are mounted at your own risk!
Repair must be carried out by a specialist for refrigerating devices!
Observe safety regulations of your country!
Cycle program
START Program 7
This program is used for a continuous test (e.g. factory test) of the centrifuge. Always 3 runs with a
subsequent interval of 1 min are carried out. After the 4th run the interval is 10 min. The FC is switched
off after each run.
The nominal values of the run can be changed after the start of the service program (pre-setting:
nominal rotational speed of the rotor, 10 min, 4 °C).
Display of the pre-set nominal values, it is possible to change them
Start cycle run
Display after each run
xx interval time in min/s
An individual run can be aborted by means of the STOP button. The lid can be opened during the
interval.
You can only quit the cycle program by switching off the device.
4 0 4000 10
PAUSE x x
MSTM/0408