Centrifuge 5804/5804 R and 5810/5810 R
4 Disassembly / Assembly
4-10 Disassembly/Assembly
Retrofitting rotors
Retrofitting must only be carried out by a trained service technician.
An imbalance adjustment is required after retrofitting the EPROM.
Retrofitting without installation is possible from the following serial numbers onwards:
For Rotor A-4-81/MTP/Flex please note:
5810R:
The 5810 R can only be upgraded from serial no. 660.
Rotor Software Centrifuges / Serial numbers
5804 5804 R 5810 5810 R 5804 5804 R 5810 5810 R
230 V 230 V 230 V 230 V 120 V 120 V 120 V 120 V
T-60-11 3.0 1380 1796 1100 2925 1410 1820 990 2949
A-2-DWP up to 3,600 1/
min
3.0 1380 1796 1100 2925 1410 1820 990 2949
A-2-DWP up to 3,700 1/
min
3.3 1980 2421 1680 4305 1970 2493 1690 4461
A-4-81 / MTP / Flex 3.5 / / 2162 6191 / / 2186 6203
F-45-48-PCR 3.7 2904 3897 2306 6743 2904 3861 2366 6743
5810 R Serial no. Frequency converter Order no.
230 Volt > 660 and < 1220 Frequency converter with extended
temperature range required
5810 855.250
120 Volt > 660 and < 1180 Frequency converter with extended
temperature range required
5810 855.241
5810 Serial no. Frequency converter Order no.
230 Volt < 2162 NEW frequency converter required 5810 855.250
120 Volt < 579 Frequency converter with extended tempera-
ture range required
5810 855.241
MSTM/0408