27
Installation
Concentrator plus/Vacufuge® plus
English (EN)
4.2.5 Checking the delivery
1. Check that the delivery is complete (see Delivery package on p. 11).
2. Check all parts for any transport damage. Contact your dealer if any damage is found.
4.3 Installing the instrument
4.3.1 General installation
Abb. 4-2: Connection options of the complete system
Fig. 4-2: Connection options of the complete system
To pack the device, perform the steps in reverse order.
WARNING! Risk from incorrect supply voltage
Only connect the device to voltage sources which correspond to the electrical
requirements on the name plate.
Only use sockets with a protective earth (PE) conductor and suitable power cable.
WARNING! Risk to health from escaping substances.
No vapors from toxic liquids and pathogenic germs must escape.
Ensure the required condensation and separation of vapors using suitable cold traps or
chemical traps.
Please note the personal protective equipment (gloves, clothing, goggles, etc.), extraction
and safety class required for the laboratory.
Select a setup for the collection and/or discharge of liquids and gases which corresponds to
the applicable legal requirements and regulations for your application.