41
8 Short instructions
It is essential to read the operating manual completely before working with the short instructions.
Switching on the device
Programming
Select the FILES menu using the keys and press .
Changing the "Standard" example program
Select Standard using the key and press .
Select the program commands using the key, enter the settings or the texts and confirm by pressing .
Using the , , , keys, move the cursor to or between the entry positions.
Program commands
T Temperatures, times, increments, ramp, gradient
HOLD To maintain temperature entered until program is continued with
Enter
(4–99
°
C)
PAUSE To interrupt program (e.g. for addition of reagent) until program is continued with
Enter
(0–11 characters)
GOTO/REP To repeat program parts and number of cycles (1–40 lines, 1–99 repeats)
SOUND Acoustic signal (1–10 tones)
LINK To branch to other programs (1–5 programs)
Confirm entries by pressing .
Enter the options for the temperature command using the key. The cursor must be positioned on the program line
number.
Complete programming by pressing .
Save the program.
CNTRL To define temperature regulation
BLOCK For the thermoblock
TUBE For different tubes
Lid To determine the lid temperature
NOWAIT To start a program immediately
WAIT Program waits for lid temperature
AUTO Lid switches off at < 22
°
C
and time > 5 min.
FIX Lid switches off at end of program
T Temperature (0–99
°
),
Time (0:00:01–9.59:59)
Temperature increment (
±
0.0 –
±
10.0
°
),
Time increment (0:01–1:00),
R Ramp (0.3–3
°
C/s),
Ramp increment (
±
0.1 –
±
1.0
°
/s)
G Gradient (
±
10
°
C)
(for Mastercycler
®
gradient only)
Menu name
To start any program
Programming
System settings
To preheat the lid
To temperature-control block and heated lid
Program in processing level
Nominal/actual block temperature
Lid temperature
(only when lid is switched on)
Main-Menu
Start
FILES . .
OPTIONS . .
Lid
Incubate
STANDARD
– –/25.0° Lid:105° 10:22:06
/
To edit the program in the processing level
To load and process the example program
To delete any program
To load and edit any program
To create a new program
FILES . .
Edit
Load
Standard
New
Delete
STANDARD
– –/22.0° 10:22:06
FILES . . Edit STANDARD
Edit CNTRL BLOCK
Load LID=105°
Standard NOWAIT AUTO
New 1 T=94.0° 0:02:00
Delete 2 T=94.0° 0:00:15
STANDARD 3 T=44.0° 0:00:15
– –/22.0° 10:22:06
FILES . . STANDARD
Edit 3 T=44.0° 0:00:15
Load 4 T=72.0° 0:00:30
Standard +0.0° +0:01
New R= 3.0°/s +0.0°/s
Delete G=0.0°
STANDARD 5 GOTO
2 REP 24
– –/22.0° 10:22:06
8 Short instructions
Time