52
12 Troubleshooting
M
emory Error The internal memory for programs
is defective.
Evaluate further messages.
M
emory full! Memory is full. Before storing, delete programs
which are not required.
M
emoryCard
(Blank)
Card has incorrect logical
identification and/or is blank.
Card must be formatted.
Automatic procedure when program
is saved.
M
emoryCard
Defective!
Card is defective. Use a new card.
M
emoryCard
HdChksumErr
Error in check sum. Card must be formatted.
Automatic procedure when program
is saved.
M
emoryCard
ManIDErr
Incorrect identification. Card must be formatted.
Automatic procedure when program
is saved.
M
emoryCard
MemOverErr!
Incorrect memory address. Card must be formatted.
Automatic procedure when program
is saved.
M
emoryCard
MemRangeErr
Error in memory range. Card must be formatted.
Automatic procedure when program
is saved.
M
emoryCard
MemSizeErr
Incorrect memory size. Card must be formatted.
Automatic procedure when program
is saved.
M
emoryCard
MemUnderErr!
Incorrect memory address. Card must be formatted.
Automatic procedure when program
is saved.
M
emoryCard
Missing!
No card has been inserted. Insert card.
M
emoryCard
Not Valid
Card is invalid. Insert card correctly
(check that card is facing in the
correct direction).
Card should be formatted before
saving begins.
M
emoryCard
Read Err!
Card cannot be read. Insert card correctly
(check that card is facing in the
correct direction).
Error message Cause Solution