54
12 Troubleshooting
V
alue out of range! Value is not within the permitted
value range.
The closest permitted value is
entered automatically.
W
arn Ramp Max
...
Line ...
The ramp is limited to 3
°
C/s. Check entries for ramp and ramp
increment, and change entries
if necessary.
W
arn Ramp Min
...
Line ...
The ramp is limited to 0.3
°
C/s. Check entries for ramp and ramp
increment, and change entries
if necessary.
W
arn Temp Max Temperature limits of 99
°
C
exceeded. If program is started,
temperature is limited to 99
°
C.
Check temperature commands
and reduce value of temperature
increment so that only values below
99
°
C can be reached.
W
arn Temp Min Temperature limit of 4
°
C not
reached. If program is started,
temperature is limited to 4
°
C.
Check temperature commands
and modify values of temperature
increment so that only values above
4
°
C can be reached.
W
arn Time Max Cycle time of 9:59:59 exceeded.
If program is started, cycle time is
limited to 9:59:59.
Check temperature commands and
modify values of time increment so
that only values below 9:59:59 can
be reached.
W
arn Time Min Cycle time of 0:00:01 not reached.
If program is started, cycle time is
limited to 0:00:01.
Check temperature commands and
modify values of time increment so
that only values above 0:00:01 can
be reached.
Error message Cause Solution