4 Technical Data
16 / 62
80116-276_V24169
2022-01
Residual volume display VIO display
Residual pressure display Pressure gauge
The APC 2 switches off when the input pressure is < 3 x 10
5
Pa < 3 bar < 43.5 psi
Dimensions and weight
Width x height x depth 410 x 80 x 370 mm
Weight 4.8 kg
Ambient conditions for transport and storage of unit
Temperature -40 °C to + 70 °C
Relative humidity 10% – 95%
Ambient conditions for operation of unit
Temperature +10 °C to + 40 °C
Relative humidity 15% – 80%, noncondensing
Acclimatizing
If the unit has been stored or transported at temperatures below +10 °C or above +40 °C, the unit will
require approx. 3
hours to acclimatize at room temperature.
Standards
Classification according to EC Directive 93/42/EEC IIb
Type as per EN 60 601-1 CF