10 Maintenance and servicing
38 / 62
80116-276_V24169
2022-01
Technical safety check – step by step
For simplification the device to be tested is referred to below as the "test
specimen".
Safety information
ATTENTION! In the event of a fault occurring in the test specimen or individual com-
ponents during the technical safety check the test steps taken so far no
longer apply.
Remedy the defect and repeat the technical safety check from the
beginning.
IMPORTANT! For all tests the test specimen must be connected to a VIO HF surgical
unit and be connected to the same via ECB. In the case of tests con
-
cerning grounded conductor resistance, ground leakage current, and
patient leakage current it is always the device combination of VIO HF
surgical unit and the actual test specimen (APC
2) which is tested
because the test specimen does not have it own built-in power supply
unit.
IMPORTANT! If the test specimen is also provided with an HF socket in addition to
the APC socket, its function must be verified in accordance with the
Safety Check instructions for the VIO electrosurgical unit connected to
the test specimen.
IMPORTANT! If the test specimen is provided with two APC sockets, operation of both
sockets must be verified in accordance with the Safety Check instruc
-
tions.
IMPORTANT! It is assumed that the user knows how to operate the test specimen,
the test equipment, the measuring equipment, and auxiliary test equip
-
ment. The test instructions only apply in conjunction with the relevant
test steps.
IMPORTANT! Test equipment, measuring equipment, and auxiliary test equipment
(cables, test boxes, etc.) are listed separately at the beginning of each
test unit. Where Erbe article numbers are specified, only original Erbe
test equipment, measuring equipment, and auxiliary test equipment
may be used.
IMPORTANT! For this test only Erbe pressure reducers may be used. For article num-
bers see chapter "Testing and measuring equipment".
IMPORTANT! The test report for the technical safety check can be requested from
Erbe Technical Service Tübingen. For the address see address sheet on
last page.

Table of Contents

Other manuals for Erbe APC 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Erbe APC 2 and is the answer not in the manual?

Erbe APC 2 Specifications

General IconGeneral
BrandErbe
ModelAPC 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals