43 / 62
10 • Maintenance and servicing
80116-276_V24169
2022-01
No leaks at 5 bar input pressure
Testing and measuring equipment
Test setup
• The test specimen is connected to the power supply via a VIO electro-
surgical unit.
• The argon gas bottle is connected to the test specimen.
• With APC socket: The pressure gauge is connected to the gas output of
the APC socket.
• With FiAPC socket: The Test adapter FiAPC is connected to the FiAPC
socket. The pressure gauge is connected to the gas output of the test
adapter.
Test procedure
1. Open the argon gas bottle valve.
2. Call up "TP valves (APC)" test program.
IMPORTANT! For this test only Erbe pressure reducers may be used. For article num-
bers see chapter "Testing and measuring equipment".
Erbe Art. No. Description
20100-108 Manometer
20134-XXX
1
1. Depending on the bottle connection.
Pressure regulator
– Argon gas bottle
20100-175 FiAPC test adapter (only if a FiAPC socket is
available)