10 • Maintenance and servicing
52 / 62
80116-276_V24169
2022-01
Test on APC socket (Activation
CUT/COAG)
Test setup
• The test specimen is connected to the power supply via a VIO S HF sur-
gical unit.
• The Basic Program is selected.
• The NE patient cable with the shorted adapter cable is connected to
the NE socket of the VIO S.
Test procedure
1. Plug the MF connector of the APC handle (20132-043) into the test
specimen.
2. Press CUT button on the APC handpiece. On the test specimen CUT
must be activated.
3. Press COAG button on the APC handpiece. On the test specimen COAG
must be activated.
4. Remove the APC handpiece.
1. Plug the MF connector of the VIO APC handle (20132-200) into the
test specimen.
2. Press CUT button on the VIO APC handpiece. On the test specimen
CUT must be activated.
3. Press COAG button on the VIO APC handpiece. On the test specimen
COAG must be activated.
4. Remove the VIO APC handpiece.
Test on APC and FiAPC sockets
(Activation FORCED APC)
Test setup
• The test specimen is connected to the power supply via a VIO S HF sur-
gical unit.
• The argon gas bottle is connected to the test specimen.
• The NE socket of the VIO S is connected to one face of a metal panel.
On the metal panel there is a viscose sponge soaked with a 0.9 %
saline solution.
• With APC socket: The VIO APC handpiece with applicator is plugged
into the test specimen.
• With FiAPC socket: The FiAPC probe is connected to the FiAPC socket.
The dual-pedal footswitch with ReMode is connected to the VIO S.
IMPORTANT! For this test only Erbe pressure reducers may be used. For article num-
bers see chapter "Testing and measuring equipment".