Polymerase Chain Reaction Cabinets
34
Problem 6 Blank LCD
PROBLEM
POSSIBLE CAUSE
CORRECTIVE ACTION
Blank LCD
Connection problem
Unplug the cabinet from the main supply.
Open the top access cover and locate electrical panel
behind it.
Open electrical panel and see Layout A at the end of this
section to locate the main board inside electrical panel.
Check whether LCD FRC (Flat Ribbon Cable) has been
inserted properly into its socket on the main board (see Fig.
6.1).
Fig 6.1