8. ELECTROMAGNETIC IMMUNITY
Guidance and manufacturer's declaration – electromagnetic immunity
CSPOLA600 is intended for use in the electromagnetic environment specified below. The customer or the
user of CSPOLA600 should assure that they are used in such an environment.
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical Fast
Transient/burst IEC
61000-4-4
±2 kV for power
supply lines
±1 kV for I/O lines
±2 kV for power
supply lines
±1 kV for I/O lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Voltage dips, short
interruptions and voltage
variations on power
supply input lines IEC
61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
) for
5 cycles
70 % U
T
(30 % dip in U
T
) for
25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
<5 % U
T
(>95 % dip in U
T
) for
0.5 cycle
40 % U
T
(60 % dip in U
T
) for
5 cycles
70 % U
T
(30 % dip in U
T
) for
25 cycles
<5 % U
T
(>95 % dip in U
T
) for
5 s
Mains power quality should be that of a
typical commercial or hospital
environment.
If the user of the CSPOLA600 requires
continued operation during power mains
interruptions, it is recommended that the
CSPOLA600 be powered from an
uninterruptible power ssupply or battery.
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Note: U
T
is the AC mains voltage prior to application of the test level.
USER MANUAL> II. SAFETY INFORMATION
17
CSPOLA600 - Polarized visual chart system > V5 - 03-2018