7-2 Error message RETRY appears
Main causes when error message RETRY appears:
When the image is not caught accurately because of examinees condition (blinking or not
looking at target etc.)
Alignment is not achieved correctly
Failure of control board
Failure of optical unit
Can a model eye be measured?
NO
YES
Check the condition of subjects eye.
Evaluate the Freeze Image.
Refer to 11
RETRY is displayed.
Realign correctly and take
measurements again.
Refer to ‘6.2 Alignment’ of
operation manual
Measurement cannot be taken in
case of disorder such as cataracts etc.
If the pupil is smaller than reticle mark,
measurements might not be taken.
Darken the surround and try to take
measurements again.
Abnormal case
Abnormal case
Normal
OK