41
2) Make sure that the cursor is positioned at the
DPT_A.
(Usually the cursor is at this position but if it is not,
move the cursor to the DPT_A with DOWN switch.)
3) Input the indicated value of the model eye in DPT_A by pressing + or .
(Input -0.100 in case that the indicated value of the model eye is S: -0.1D.)
4) Align with the mode eye equivalent to 0D (align in
such a way that the Kerato ring becomes the thinnest)
and press the measurement start switch. After 10
times of continuous measurement, the values are
input in AXS_A, CYL_A and SPH_A automatically.
5 Press Exit switch. The screen returns to the selection screen of REF CALIBRATION
and the calibration contents (values) are stored.
This is the end of calibration of the model eye equivalent to 0D.
When conducting calibration of the model eyes equivalent -10D, 10D and -5D, select
MODEL EYE B to F and follow the same procedures.
If the alignment moves during a measurement, the accuracy of the input
values is lowered. Prevent the alignment from moving during a measurement.
If turning off the power without pressing Exit, the contents of calibration
are not stored. Make sure to return to the REF CALIBRATION screen.

Other manuals for Essilor AKR 550

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Essilor AKR 550 and is the answer not in the manual?

Essilor AKR 550 Specifications

General IconGeneral
BrandEssilor
ModelAKR 550
CategoryMedical Equipment
LanguageEnglish

Related product manuals