Instruction Manual COND 6+, TDS 6+, SALT 6+
-19-
Entering Advanced Setup Mode
Cell Constant. Select k = 0.1, 1.0, or 10.0.
Default value is 1.0. (COND 6+, TDS 6+ only)
Select Automatic Calibration. “Yes” for auto calibration
and “No” for manual calibration.
Default value is “Yes”. (COND 6+ only)
%
Adjust Temperature Coefficient from 0.0 to 3.0 % per °C.
Default value is 2.1 % per °C.
C°
Normalization Temperature. Select “20 °C” or “25 °C”.
Default value is 25 °C.
Adjust TDS factor from 0.4 to 1.0.
Default value is 0.5. (TDS 6+ only)
Select Single Point Calibration. Select “Yes” or “No”.
Default value is “Yes”.
Select Measurement Unit, TDS (ppt) or percentage (%).
(SALT 6+ only)
User reset to factory defaults. Select “Yes” or “No”.
Default value is “no”.
Overview of Advanced Setup