EasyManua.ls Logo

EUTECH INSTRUMENTS COND 6 PLUS - Normalization Temperature; Single-Point Calibration (COND 6+, TDS 6+ Only)

EUTECH INSTRUMENTS COND 6 PLUS
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instruction Manual COND 6+, TDS 6+, SALT 6+
-22-
5.6 Normalization Temperature
You can set the meter to normalize its measurements to a temperature of either 25 °C
or 20 °C. Default value is 25 °C.
1. Enter advanced setup as described in
Section 5.1.
2. Press or until [t.nr °C] appears. Press ENTER.
3. Press or to select [25.0 °C] or [20.0 °C].
4. Press ENTER to select and return to the [t.nr °C]
setup function.
5. Press or to move to the next setup function or
press CAL to exit to measurement mode.
5.7 Single-Point Calibration (COND 6+, TDS 6+ only)
Select “YES” to apply a single calibration value across all ranges.
Select “no” to allow separate calibrations for each range, or
to restrict an individual calibration so that it is applied to one
range only. Default value is “Yes”.
1. Enter advanced setup as described in Section 5.1.
1. Press or until [S.P.CA] appears. Press ENTER.
1. Press or to select [Yes] or [no].
1. Press ENTER to select and return to the [S.P.CA]
setup function.
1. Press or to move to the next setup function
menu or press CAL to exit to measurement mode.
C°
C°
C°
HOLD
ENTER
HOLD
ENTER

Related product manuals