EXCELL PRECISION CO., LTD
3-7 Digital calibration
Set CAL switch to ON
Select Digital calibration
Press the Key
Two sec. later
q
Press the key
Two sec. later
Input the span voltage
Press the key
Two sec. later
Enter the weighing capacity
Press the key
Calibration completed set calibration switch to the OFF position
Please refer to the error message list if the display shows . X .
Example:
Zero Voltage 0.00036 mV/V(incl. dead load)
Span Voltage 2.90000 mV/V
Capacity 30 000 divisions
Increment flashing digit
Decrement flashing digit
Move flashing point left
Move flashing point right
Store data in memory
Exit / Escape
Zero voltage calibration.
Method 1
Input zero voltage
With no weight on the platform or in the hopper
press the key to set zero.
Display shows nothing
after calibration