Appendix 5 Examples for Input and Output of Modbus
Examples here use Appendix 3 and can also be applied to appendix 4, with different
Modbus Data Address.
CRC : Error check code
Function code 03: Modbus (40000~40001) Weight Display
Register's starting
address is
0 = 00 00
(HEX)
1 register per 2 bytes,
and total of 4 bytes
Data in 40000
500
(10)
= 01 F4
(HEX)
40000~40001,
2 registers are used
Function code 05: Modbus (01000) write zero command
1000
(10)
= 03 E8
(HEX)
this address for "zero
on/off"
1000
(10)
= 03 E8
(HEX)
this address for "zero
on/off"
FF 00 :zero on
00 00 : zero off
FF 00 :zero on
00 00 : zero off
Function code 06: Modbus (41004) write zero band value
address for zero
band value
1004
(10)
= 03 EC
(HEX)
address for zero
band value
1004
(10)
= 03 EC
(HEX)