Appendix 2 Function Table
Specification Calibration
Zero range = calibration zero point (Max.
capacitysetting value %)
Time and range of zero tracking should be
use at the same time. If the time is set to 0.0,
the zero tracking function is disabled
Range of zero tracking = (setting value½)D ,
D=min. division
Range and time of zero tracking should be
use at the same time. If the range is set to 0,
the zero tracking function is disabled
Investigate
time in stable
Investigate time and range should be use at
the same time. If the time is set to 0.0, the
investigate time is disabled
Investigate
range in stable
Investigate time and range should be use at
the same time. If the range is set to 0, the
investigate range is disabled
Weight
unstable,
function ZERO
and TARE
Gross Weight
is negative,
function TARE