7 CO210-en-EU_V1.2 8/14
Calibration
CO2 CALIBRATION
The meter is calibrated to a standard 400ppm CO2 concentration at the factory
NOTE: When the accuracy becomes a concern or after a year of use, return the meter to Extech for
calibration.
CAUTION: Do not calibrate the meter in an atmosphere of unknown CO2 concentration.
ABC (Automatic Baseline Calibration)
ABC (Automatic Baseline Calibration) establishes a baseline calibration to eliminate the zero drift of
the infrared sensor. The ABC function is always “ON” when the meter is turned on. ABC is designed
to calibrate the meter at the minimum CO2 reading detected during 7.5 days continuous monitoring
(power on). It assumes that the area being tested receives fresh air with a CO2 level of
approximately 400ppm at some period of time during the seven days. It is not suitable to use
desktop CO2 in closed areas with consistently high CO2 levels 24 hours a day.
Manual Calibration
Manual calibration is suggested to be done outdoors on a sunny day with good ventilation and fresh
air where the CO2 level is approximately 400 ppm. Do not calibrate on a rainy day because high
humidity will affect the CO2 level in air.
Do not calibrate in places crowded with people or close to where high CO2 concentrations may exist
such as ventilating outlets or fireplaces.
Place the meter in the calibration site. Turn on the meter and hold down the SET, and
MIN/MAX buttons simultaneously for more than 1 second to enter CO2 calibration mode.
“400ppm” and “CO2” will blink while the calibration is in process.
The calibration will take about 30 minutes. When the calibration is complete, the blinking stops and
the meter returns to normal operation. To abort the calibration, press the RESET button for more
than 1 second.
Note: Keep away from any animal, human or plant which might affect the CO2 concentration during
the calibration.
Software
The Windows based pc software provided is used to set-up the datalogger, download data and to
view the data in graphical or text format. After installing the software, view the HELP file for
operational information.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Other manuals for Extech Instruments CO210

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Extech Instruments CO210 and is the answer not in the manual?

Extech Instruments CO210 Specifications

General IconGeneral
BrandExtech Instruments
ModelCO210
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals