6 CO240-en-US_v1.9 10/22
Maintenance and Troubleshooting
Battery Replacement
When the battery icon appears on the LCD, the batteries must be replaced.
In a downward motion, slide the rear battery compartment cover off of the meter housing.
Remove the old batteries and replace with new batteries (all of the same type) observing correct
polarity. This meter uses six (6) 1.5V AAA batteries. Please replace the battery compartment
cover before attempting to use the meter.
Battery Safety Reminders
o Please dispose of batteries responsibly; observe all relevant regulations.
o Never dispose of batteries in a fire; batteries may explode or leak.
o Never mix battery types; install new batteries of the same type.
Troubleshooting / Error Message Displays
Erratic Readings Replace the batteries or use the AC/DC adaptor and check again.
ER1 Display CO
2,
Temperature or Humidity sensor failure or a sensor communication
error. Send for service if display persists.
ER2 Display Out-of-range condition for CO
2,
Temperature or Relative Humidity.
Detected reading exceeds CO
2,
Temperature or Humidity range
specification. This error is normal when clearing the maximum and
minimum readings in the MAX/MIN mode.
“9999” Display Reading exceeds the CO
2
maximum measurement range or the battery
power is weak.