6 EC170-EU-EN V1.1 7/13
MeasurementPreparations,Notes,andConsiderations
Accuracyisgivenas%fullscale;thereforeusingthelowestrangewillyieldthebest
accuracy.
Themeter’sdisplaywillindicateE02orE03ifthemeasuredvalueisbelow(E02)orabove
(E03)specifiedlimitsofthemeter.Ifthisoccurs,pleaseselectanotherrangeasdescribed
in
theManualRangediscussioninpreviousparagraph.
Thetemperaturecoefficientisfixedandcannotbeadjusted.
Thenormalization(reference)temperatureisfixed(25
o
C);thenominalvalueiscorrectfor
mostapplications.
Rinsetheprobewithdeionizedordistilledwaterbeforeusetoremoveimpuritiesthatmay
adheretotheelectrode.Whenthemeterhasbeenidleforalongperiod,soaktheelectrode
foratleast30minutesbeforeuse.
Whendippingtheprobeintoasamplesolution,besuretoeliminateairbubblestrappedin
theprobe’sslot.Toremoveairbubbles,givetheprobeagentlestirwhilesubmergedinthe
solution.
Whentakingameasurement,stirtheprobegentlyinthesampletocreateahomogenous
sample.
Allowafewsecondstoelapsefortheprobeandthesampletoreachtemperature
equilibrium.Ideally,wait15minutestoachievemaximumaccuracyandbesttemperature
compensation.
Theunitofmeasureiconwillflashonthemeter’sdisplaywhilestabilizationistakingplacein
measurementmode.Whenstabilizationisachieved,themeter’siconwillstopflashing.
PresstheHOLDbutton tofreezeadisplayedreading.Pressagaintoreleasethedisplay.
SalinityMeasurement
1. ReadtheMeasurementPreparationsectionabovebeforecontinuing.
2. Inserttheelectrodeintothesampleensuringthattheelectrodesarecompletelysubmersed.
3. Slowlystirthesolutionwiththeelectrodetoremoveairbubbles.
4. ThemeterusestheAutorangeutilitytofindtheproperrangeandthendisplays
thesalinity
readingatthecenteroftheLCDwiththetemperaturereadingshownonthelowerLCDarea
insmallerdigits.
FinishingaMeasurementSession
Afterameasurementsession:
Rinsetheelectrodeindeionizedordistilledwaterandstoredry.
Affixtheprotectivecapovertheelectrodewhenstoring.
Iftheunitistobeleftununusedforlongperiods,removeandstorethebatteriesseparately.
AutomaticPowerOFF(Sleepmode)
ThemeterwillautomaticallyswitchOFFafter20minutesofinactivity.TodisabletheSleep
Mode:Withtheinstrumentswitchedoff,pressandholdtheSETandHLD/CALbuttons
simultaneouslyuntilthe‘n’iconappearsonthedisplay.Releasethebuttonsandthemeterwill
powerup.ThemeterwillnowstayswitchedONuntiltheusermanuallyswitchesitOFF.The
meterrevertstothe‘SleepModeactive’stateeachtimeitispowereddown.