13 Doc. No. 06EN003367 Rev. E
ST50 Mass Flow Meter Fluid Components International LLC
Use the V menu to set up the 4-20 mA analog outputs (including NAMUR configuration) and source/sink (pulse) outputs.
Note: The display comes up to the last setting saved and stays for 2 seconds. If N or [ENTER] is entered, the menu
proceeds to the Pulse out. If Y is entered, the display moves to the selection options and/or asks for confirmation. If you
miss the option, select [Enter] repeatedly to loop around
“V” Menu Output Configuration Setup
Example: COMMAND V (Reference Table 3)
Case: 4-20 mA #1 = Flow, 4-20 mA #2 = Temperature, NAMUR = Low, Source Out = Pulse, Sink = Alarm1
Pressing [V] [ENTER] will display Output Mode Selected” :
The last saved mode of the next menu item displays:
“4-20 mA #1 = Flow” “ 4-20 mA #2 = Temp” ...followed by the prompt:
“Change? (Y/N)”
Press [ENTER] (no change).
The last saved mode of the next menu item displays:
“NAMUR: Off” ...followed by the prompt:
“Change? (Y/N)” Select Y [Enter]. The display reads,
“NAMUR: Off” ...followed by the prompt:
“Enter 1 to make the selection #.” Select [ENTER] Bypassing “1” the next display reads,
“NAMUR: Low” ...followed by the prompt:
“Enter 2 to make the selection #.” Select 2 and [ENTER] “2” sets the new NAMUR configuration.
The last saved mode of the next menu item displays:
“Source: Pulse” “Sink: Pulse” ...followed by the prompt:
“Change? (Y/N)” Select Y [Enter]. The display reads,
“Source: Pulse” “Sink: Pulse” ...followed by the prompt:
“Enter 1 to make the selection #.” Select [ENTER]. Bypassing “1” the next display reads,
“Source: Pulse” “Sink: Alarm1” ...followed by the prompt:
“Enter 2 to make the selection #.” Select 2 and [ENTER]. “2” sets the new source/sink configuration.
The last saved mode of the next menu item displays:
“PFactor: 1.000” ...followed by the prompt:
“Change? (Y/N)>” Respond with “Y” to enter a factor anywhere from 0.001
to 1000. A pulse factor of 1.000 outputs 1 pulse per unit
of flow. If no change, select N and/or [ENTER] to
continue.