Fluid Components International LLC
This page is subject to proprietary rights statement on last page
14 Doc. No. 06EN003367 Rev. E
ST50 Mass Flow Meter
The last saved mode of the next menu item displays:
“Sample Period” ...followed by the prompt:
“Change? (Y/N)>” Respond with “Y” to enter a sample period value
from 0.5 to 5 seconds.
If no change, select N and/or [ENTER] to continue.
The last saved mode of the next menu item displays:
Source state: “High to Low” ...followed by the prompt:
Change to “Low to High?>” Respond with “Y” to toggle to the alternate setting.
If no change, select N and/or [ENTER] to continue.
The last saved mode of the next menu item displays:
“Switchpt1” “0.0000000” ... followed by the prompt:
“Change? (Y/N)>” Respond with “Y” to enter a setpoint value (value is
in same units as the flow and must be within the
calibrated range).
If no change, select N and/or [ENTER] to continue.
The last saved mode of the next menu item displays:
Sink state: “High to Low” ...followed by the prompt:
“Change to “Low to High?>” Respond with “Y” to toggle to the alternate setting.
If no change, select N and/or [ENTER] to continue to normal operation (programming finished).