Page #1 background imageLoading...
Page #1 background image
11
Fieldpiece
Job Link® System
Power
Clamp Meter
OPERATOR'S
MANUAL
Model SC680
+
INRUSH
Question and Answer IconNeed help?

Do you have a question about the Fieldpiece SC680 and is the answer not in the manual?

Fieldpiece SC680 Specifications

General IconGeneral
DisplayLCD
ContinuityYes
Diode TestYes
Warranty1 Year
Temperature Range-40 to 752°F
Battery9V
Electrical MeasurementVoltage, Current, Resistance, Capacitance, Frequency, Temperature

Summary

Safety Information

General Safety Precautions

Use meter as specified, inspect for damage, replace leads, avoid abnormal operation.

Environmental and Measurement Safety

Avoid storms, explosive environments, adhere to CAT ratings, voltage limits.

CAUTION and Warnings

Voltage Hazards and Personal Protective Equipment (PPE)

30VAC/60VDC pose shock hazard. Use safety glasses, gloves, boots, mats.

Pre-Use Checks and Measurement Best Practices

Perform continuity test, 3-point safety method, use proper terminals, avoid grounding.

Specific Operational Safety Warnings

Disconnect leads before opening, test NCV on live wire, respect voltage limits.

Product Description

Key Features and Applications

Wireless, system efficiency, mini-splits, advanced measurements, True RMS, Inrush.

Included Accessories and Quick Start

Lists meter accessories and provides basic operating instructions.

Controls and Operation

Measurements Dial Functions

Explains each position on the rotary selector dial for measurements.

Meter Buttons Explained

Details the function of each button on the meter.

Display and Readability

Display Icons and Unit Abbreviations

Reference for symbols, icons, and abbreviations used on the display.

Easy-to-Read Display Features

Highlights the large dual display and blue backlight for clarity.

Measurement Functions - AC/DC Voltage

Volts AC (VAC) True RMS

How to measure AC voltage, ranges, accuracy, and overload protection.

Volts DC (VDC)

How to measure DC voltage, ranges, accuracy, and overload protection.

Non-Contact Voltage (NCV)

How to detect AC voltage without contact, range, and indicators.

Measurement Functions - Current and Frequency

Amps AC (AAC) True RMS

Measuring AC current via clamp, ranges, accuracy, and overload protection.

Frequency (Hz) Through Clamp

Measuring frequency via clamp, ranges, accuracy, and current requirements.

Amps DC (ADC)

Measuring DC current via clamp, ranges, accuracy, and overload protection.

Inrush AAC Current Measurement

Capturing motor starting amp draw, activation steps, and measurement period.

Measurement Functions - MicroAmps, Frequency, Duty Cycle, Capacitance

MicroAmps DC (µADC) and Frequency (Leads)

Measuring small DC currents and frequency using test leads.

Duty Cycle (%) Measurement

Measuring the duty cycle of square wave signals, ranges, and accuracy.

Capacitance (MFD) Measurement

Testing capacitors, discharge warning, ranges, and accuracy.

Measurement Functions - Resistance, Continuity, Diode

Resistance (Ω) Measurement

Testing resistance, terminal pole checks, ranges, and accuracy.

Continuity (•) Test

Checking fuses/circuits for open/closed status, indicators, and response time.

Diode Test (+)

Testing diodes for forward/reverse bias, display, and beep indicators.

Temperature Measurement and Calibration

Dual Temperature Measurement (T1, T2, T1-T2)

Measuring temperature with Type K thermocouples and delta T.

Temperature Calibration Procedure

Steps to calibrate temperature jacks using ice water.

Power Measurement

Power Measurement Explained

Details Active, Reactive, Apparent, DC Power, and Power Factor.

Power Measurement Setup and Types

How to take power readings and the types of power measured.

Phase Rotation Test

Performing Phase Rotation Test

Step-by-step guide for testing 3-phase motor rotation direction.

Advanced Testing and Expandability

Phase Rotation Testing Tips

Important notes and conditions for phase rotation testing.

Modular Expandability with Accessory Heads

Information on compatibility with Fieldpiece Accessory Heads.

Job Link System Integration

Sending Wireless Measurements

Steps to send measurements wirelessly to the Job Link mobile app.

Wireless Specifications and Compatibility

Details wireless range, frequency, and device compatibility.

Meter Functions

Key Operational Functions

How to use Hold, Max/Min, and Auto Power Off (APO) features.

Utility and Safety Functions

High Voltage Warning, Battery Replacement, Backlight, Manual Ranging.

Specifications and Certifications

Technical Specifications

Details display, environment, power, battery life, dimensions, weight.

Safety Certifications and Standards

Lists safety certifications like UL, CE, FCC, and relevant standards.

Regulatory Compliance

FCC and IC Statements

Compliance information for FCC and Industry Canada regulations.

Warranty and Service

Limited Warranty and Obtaining Service

Details product warranty terms and how to obtain service.