13 Data logger and transmission of data FLUXUS F808, F809
UMFLUXUS_F808_8091V1-2-1EN, 2019-08-23 125
Storing of the signal amplitude
Storing of the sound speed
Storing of the diagnostic values
13.1.4 Measurement with activated data logger
• Start the measurement.
If Output Options\Store Meas.Data has been activated and Special Funct.\SYSTEM settings\
Ringbuf-
fer is deactivated, an error message will be displayed as soon as the data logger is full.
The storing will be stopped.
13.1.5 Deleting measured values
13.1.6 Available data logger capacity
If the data logger is empty and a measurement is started with one physical quantity on one measuring channel without
storing the totalizer and further values, approx. 100 000 measured values can be stored. The max. available data logger
capacity can be displayed:
Max. 100 series of measured values can be stored. The number of series of measured values depends on the total num-
ber of measured values stored in the previous series of measured values.
If on is selected and the data logger is activated, the amplitude of the measured signal will
be stored together with the measured values. Press ENTER.
If on is selected and the data logger is activated, the sound speed of the fluid will be stored
together with the measured values. Press ENTER.
If on is selected and the data logger is activated, the diagnostic values will be stored to-
gether with the measured values. Press ENTER.
Enter the measuring point number. Press ENTER.
If arrows are displayed in the lower line on the right, ASCII text can be entered. If digits are
displayed, only digits, point and hyphen can be entered.
For the setting of the input mode see section 15.2.3.
Press ENTER.
The error message will be displayed periodically.
Select Special Funct.\Delete Meas.Val. Press ENTER.
Select yes or no. Press ENTER.
Select Special Funct.\Instrum. Inform. Press ENTER.
The type and the serial number of the transmitter will be displayed in the upper line.
The max. available data logger capacity will be displayed in the lower line (here: 18 327 ad-
ditional measured values can be stored). Press key BRK twice to return to the main menu.
Store Amplitude
off >ON<
Store c-Medium
off >ON<
Store diagnostic
off >ON<
A:Meas.Point No.:
xxx (↑↓← →)
DATA MEMORY
OVERFLOW!
Special Funct. ↕
Delete Meas.Val.
Really Delete?
no >YES<
Special Funct. ↕
Instrum. Inform.
F80X -XXXXXXXX
Free: 18327