5 Selection of the Measuring Point FLUXUS G80x
UMFLUXUS_G8V4-6-1EN, 2018-10-10 21
disturbance source: T piece
supply line: l 100 D return line: l 20 D
disturbance source: diffuser
supply line: l 60 D return line: l 10 D
disturbance source: valve
supply line: l 80 D return line: l 20 D
disturbance source: reducer
supply line: l 20 D return line: l 10 D
disturbance source: compressor
supply line: l 100 D
Tab. 5.2: Recommended distance from disturbance sources; D – nominal pipe diameter at the measuring point, l – recommended distance