7 Technical specifications
Table 7.3 Peak hold: peak maximum/peak minimum (AC only, Non TRMS)
Function Range Accuracy
ACV
140.0 V
±(3.0% + 15 dgt)
1400 V
ACA
140.0 A
±(3.5% + 15 dgt)
850 A
Overload protection: 1000 V
rms
, 600 A
rms
Peak Hold Response time: 200μs
Accuracy defined for sine waves, ACV > 5 V
rms
/ACA ≧ 5 A
rms
, frequency 50–400
Hz. Only suitable for repetitive events.
Table 7.4 Frequency
Function Range Accuracy
Frequency
20.00–99.99 Hz
±(0.5% + 3 dgt)
20.0–999.9 Hz
0.020–9.999 kHz
Overload protection: 1000 V
rms
, 600 A
rms
.
Sensitivity:
• 10–100 V
rms
for AC 100 V range.
• 10–100 A
rms
for AC 100 A range (>400 Hz unspecified).
• 100–1000 V
rms
for AC 1000 V range.
• 100–600 A
rms
for AC 600 A range (>400 Hz unspecified).
The reading will be 0.0 for signals below 10.0 Hz.
Table 7.5 Total harmonic distortion
Function Range Accuracy
ACA/ACV
99.9%
±(3.0% + 10 dgt)
#T559825; r. AI/33557/33557; en-US 35