Handling the camera
7
7.15.2 AutoCal procedure
1. Dip the calibration target in water for 1 second and let the excess drip off.
2. Tape or hang the calibration target on a wall.
3. Mount the new lens on the camera according to the procedure in section 7.14
Changing camera lenses. When the lens is mounted, the calibration wizard starts
automatically.
#T810253; r. AL/84068/84441; en-US
33