Calibrator
Remote Commands
43
Command Description
*TST?
Initiates a series of self-tests, then returns a 0 for pass or a 1 for fail. If any faults are
detected, they are logged into the fault queue where they can be read by the ERR?
query.
*WAI
Prevents further remote command execution until all previous remote commands
have been executed.
Table 21. Error Mode Commands
Command Description
EDIT
Set the edit field. PRI is specified for the output value in single output functions and
the primary output value in dual output functions.
EDIT? Returns the edit field setting.
ERR_REF Selects the error reference source.
ERR_REF? Returns the presently selected error reference source.
ERR_UNIT Chooses how DUT error is shown.
ERR_UNIT? Returns presently selected value of ERR_UNIT.
INCR
Increments or decrements the output (as selected by the edit field) and enters error
mode, the same as if you use the output adjustment knob in local operation.
MULT Multiplies the reference magnitude (as selected by the edit field).
NEWREF
Set the reference value to be the present Calibrator output value, the same as if you
tap New Reference in local operation.
OLDREF
Set the Calibrator output to the previously-programmed reference value, the same as
if you push E in local operation.
OUT_ERR? Returns the DUT error computed after shifting the output with the INCR command.
REFOUT?
Returns the value of the reference, which is the output values of the Calibrator the last
time a new reference was established with an OUT, NEWREF, or MULT.
Table 20. Common Commands (cont.)