Reference Multimeter and 8 ½ Digit Multimeter
SCPI Commands Reference
25
Table 6. Command Summary (cont.)
Keyword Description
OFF = Secondary reading is not shown
FREQuency = Secondary reading is frequency
PERiod = Secondary reading is period
PTP = Secondary reading is peak-to-peak
PPEak = Secondary reading is positive peak
NPEak = Secondary reading is negative peak
CFACtor = Secondary reading is the crest factor
:SECondary? Returns the selected secondary reading setting
:SECondary:METHod MEASure|SINE|
SQUare|TRIangle|TSINe
Set the method by which the secondary peak-to-peak value is
calculated
MEASured = The measured positive peak minus the measured
negative peak value
SINe = The peak to peak of a sine wave calculated from the rms
value
SQUare = The peak to peak of a square wave calculated from
the rms value
TRIange = The peak to peak of a triangle waveform calculated
from the rms value
TRUNcated = The peak to peak of a truncated sine wave
calculated from the rms value
:SECondary:METHod? Returns the method by which the secondary peak-to-peak value
is calculated
[:SENSe]:CURRent:DC
:APERture <nrf>|MIN|MAX|DEF
Sets the ADC aperture value in seconds, or to the MIN, MAX, or
default setting
:APERture? [MIN|MAX|DEF]
Returns the aperture setting or the MIN, MAX, or default settings
in seconds
:APERture:MODE AUTO|FAST|MANual Sets the aperture mode
:APERture:MODE? Returns the aperture mode
:NPLC <plcs>MIN|MAX|DEF
Sets the ADC aperture in number of power line cycles or to the
MIN, MAX, or default plc setting
:NPLC? [MIN|MAX|DEF]
Returns the ADC aperture in number of power line cycles or to
the MIN, MAX, or default plc setting
:RANGe:AUTO[:STATe] <boolean> Turns Auto range ON or OFF