8588A/8558A
Remote Programmer's Manual
26
Table 6. Command Summary (cont.)
Keyword Description
:RANGe:AUTO[:STATe] ? Returns 1 for Auto range ON, 0 for auto range OFF
:RANGe <nrf>|MIN|MAX|DEF Set maximum expected value or min, max or default range
:RANGe? [MIN|MAX|DEF]
Returns the selected range or, if specified, the minimum,
maximum or default range
:RESolution <nrf>|MIN|MAX|DEF
Set maximum expected value or min, max or default resolution;
for example, range is 1 A, Resolution <nrf>. = 0.0001 (100 µA),
the measurement is returned with a resolution of +1.000E-4
:RESolution? [MIN|MAX|DEF]
Returns the selected resolution, or if specified, the minimum,
maximum or default range
[:SENSe]:DIGitize
:APERture <nrf>|MIN|MAX|DEF
Sets the ADC aperture value in seconds or to the MIN, MAX, or
default setting
:APERture? [MIN|MAX|DEF]
Returns the aperture setting or, if specified, the MIN, MAX, or
Default settings in seconds
:CURRent:COUPling AC|DC Sets the current signal coupling path to ac or dc
:CURRent:COUPling? Returns the current signal coupling path
:CURRent:RANGe
<range>|MIN|MAX|DEF
Sets the maximum expected current value or the MIN, MAX, or
Default range
:CURRent:RANGe? [MIN|MAX|DEF] Returns the current range set
:FILTer OFF|100Khz|3MHZ
Sets the digitise Filter bandwidth to OFF, 100 kHz, or 3 MHz
(default is 3 MHz)
:FILTer? Return the Digitize low pass filter setting
:VOLTage:COUPling:SIGNal
AC1M|AC10M|DC1M|DC10M|DCAuto
Set the voltage path coupling and input impedance
AC1M = AC Coupling, 1 MΩ input impedance
AC10M = AC Coupling, 10 MΩ input impedance
DC1M = DC Coupling, 1 MΩ input impedance
DC10M = DC Coupling, 10 MΩ input impedance
DCAuto = DC Coupling, maximum available input impedance
(>1 TΩ for ranges < 100 V; 10 MΩ, for the 100 V, and 1000 V
ranges)
:VOLTage:COUPling:SIGNal? Return the voltage path coupling and input impedance setting
:VOLTage:RANGe
<range>|MIN|MAX|DEF
Sets the maximum expected voltage value or the MIN, MAX, or
Default range
:VOLTage:RANGe? [MIN|MAX|DEF] Returns the voltage range set