28
Table 6. Command Summary (cont.)
Keyword Description
[:SENSe]:FRESistance
:APERture <nrf>|MIN|MAX|DEF
Sets the ADC aperture value in seconds or to the MIN, MAX, or
default setting
:APERture? [MIN|MAX|DEF]
Returns the aperture setting or the MIN, MAX, or Default
settings in seconds
:APERture:MODE AUTO|FAST|MANual Sets the aperture mode
:APERture:MODE? Returns the aperture mode
:NPLC <nrf>MIN|MAX|DEF
Sets the ADC aperture in number of power line cycles or to the
MIN, MAX, or Default plc setting
:NPLC? [MIN|MAX|DEF]
Returns the ADC aperture in number of power line cycles or to
the MIN, MAX, or Default plc setting
:MODE NORMal|TRUE|HIV Sets the 4-wire Ohms mode
:MODE? Returns the 4-wire Ohms mode setting
:LOWI[:STATe] <boolean> Sets the Low current mode ON or OFF
:LOWI? Returns the low current mode setting
:RANGe:AUTO[:STATe] <boolean> Sets Auto ranging ON or OFF
:RANGe:AUTO[:STATe] ? Returns the Auto range setting
:RANGe <range>|MIN|MAX|DEF
Sets the expected maximum input resistance, or if selected, to
the MIN, MAX, or Default setting
:RANGe? [MIN|MAX|DEF]
Returns the range setting, or if selected, the MIN, MAX, or
Default setting
:RESolution <resolution>|MIN|MAX|DEF Set maximum expected value or min, max or default resolution
:RESolution? [MIN|MAX|DEF]
Returns the selected resolution, or if specified, the MIN, MAX, or
Default resolution
:FILTer[:STATe] <boolean> Turns the input filter ON or OFF
:FILTer[:STATe]? Returns the filter setting, 0 = OFF, 1 = ON
[:SENSe]:FUNCtion <string>
(where string can be)
"CURRent[:DC]" Sets function to DCI
"CURRent:AC" Sets function to ACI
"CAPacitance" Sets function to Capacitance