101
Users Manual
22
Function Range Resolution Accuracy
Resistance
O
400.0 Ω
4.000 kΩ
40.00 kΩ
400.0 kΩ
4.000 MΩ
40.00 MΩ
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
0.5 % + 3
0.5 % + 2
0.5 % + 2
0.5 % + 2
0.5 % + 2
1.5 % + 3
Capacitance
[1]
c
50.00 nF
500.0 nF
5.000 μF
50.00 μF
500.0 μF
1000 μF
0.01 nF
0.1 nF
0.001 μF
0.01 μF
0.1 μF
1 μF
2 % + 5
2 % + 5
5 % + 5
5 % + 5
5 % + 5
5 % + 5
Frequency
[2]
Hz
(10 Hz – 100 kHz)
50.00 Hz
500.0 Hz
5.000 kHz
50.00 kHz
100.0 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 kHz
0.1 % + 3
Duty Cycle
[2]
1 % to 99 % 0.1 % 1 % typical
[3]
[1] Specifications do not include errors due to test lead capacitance and capacitance floor (may be up to 1.5 nF in the
50 nF range).
[2] All AC, Hz, and duty cycle are specified from 1 % to 100 % of range. Inputs below 1 % of range are not specified.
[3] Typical means when the frequency is at 50 Hz or 60 Hz and the duty cycle is between 10 % and 90 %.