EasyManuals Logo

Fluke 114 Calibration Information

Fluke 114
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #11 background imageLoading...
Page #11 background image
Digital Multimeters
Performance Tests
11
Table 5. DMM Performance Tests (cont)
Display Reading
Step Function Range Applied
114
[1]
115
[1]
116
[1]
117
[1]
41 VoltAlert Hi
N/A N/A N/A
Refer to
steps 1 – 5
in the
procedure
below
42 VoltAlert Lo
N/A N/A N/A
Refer to
steps 6 – 9
in the
procedure
below
[1] If using a Fluke 9100 calibrator, the Calibrator Frequency mode must be used to obtain accurate frequency.
[2] To keep from tripping the calibrator to standby, ramp up the voltage in 50 V increments with a 5 second delay between
increments.
Testing the VoltAlert Function (117 only)
Use the following procedure to verify that VoltAlert functions properly.
Note
Ensure the instrument is REMOVED from the holster prior to performing the test.
Keep the meter away from electrical noise sources during the tests, i.e., florescent
lights, dimmable lights, motors, etc.. These sources can trigger VoltAlert and
invalidate the test.
It may be necessary in steps 4 and 8 below to slightly adjust the Meter’s position for
maximum signal strength, in order to get the Meter’s beeper to sound continuously.
Refer to Figure 4 while performing the following steps.
1. Select the VoltAlert function, and verify that “Hi” is on the display. Verify that the beeper is silent and
the red LED is off.
2. Connect a double banana plug to the output voltage terminals of the calibrator (Fluke 5500A or
equivalent).
3. Set the calibrator output to 10 V at 60 Hz.
4. Hold the Meter so that the Meter’s top is vertically and horizontally centered and contacting the banana
plug’s Hi terminal. Verify that the Meter’s beeper is on continuously, and the red LED, at the top of the
display, lights up.
5. Place the calibrator in standby mode and verify that the beeper is now silent and the red LED is off.
6. Press the q(display should indicate ‘Lo’ range).
7. Set the calibrator’s output to 30 V at 60 Hz.

Other manuals for Fluke 114

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 114 and is the answer not in the manual?

Fluke 114 Specifications

General IconGeneral
Frequency range0.005 - 50 kHz
Continuity checkYes
DC voltage range0.0001 - 600 V
Resistance range1 - 600 Ω
Resistance resolution1 Ω
Resolution (DC voltage)0.0001 mV
Basic accuracy (DC voltage)1.0% + 3 (DC, 45 Hz to 500 Hz)
Temperature measurement range-40 - 400 °C
Battery type9V
Battery voltage9 V
Number of batteries supported1
Weight and Dimensions IconWeight and Dimensions
Depth46 mm
Width84 mm
Height167 mm
Weight550 g

Related product manuals