Digital Multimeters
Calibration Adjustment
15
Notes
Set the calibrator to Standby prior to changing the function switch position and after
completing adjustment of each function.
If the calibration adjustment procedure is not properly completed, the Meter will not operate
correctly.
Table 6. Calibration Adjustment Steps
Calibration Steps
Rotary
Switch
Position
114 115
[1]
116
[1]
117
[1,2]
Input
Terminals
Calibrator
Source Value
e
Ohms
N/A C01 C01 C01
[2]
No leads No leads
C01 C02 C02 C02 VΩ/+ and COM 0 V, 0 Hz
C02 C03 C03 C03 VΩ/+ and COM 300 mV, 0 Hz
C03 C04 C04 C04 VΩ/+ and COM 100 mV, 0 Hz
C04 C05 C05 C05 VΩ/+ and COM -300 mV, 0 Hz
C05 C06 C06 C06 VΩ/+ and COM 60 mV, 0 Hz
C06 C07 C07 C07 VΩ/+ and COM 600 mV, 0 Hz
m
C07 C08 C08 C08 VΩ/+ and COM 600 mV, 60 Hz
C08 C09 C09 C09 VΩ/+ and COM 600 e, 2-wire comp
C09 C10 C10 C10 VΩ/+ and COM 6 ke
C10 C11 C11 C11 VΩ/+ and COM 60 ke
C11 C12 C12 C12 VΩ/+ and COM 600 ke
C12 C13 C13 C13 VΩ/+ and COM 6 Me
[3]
C13 C14 C14 C14 VΩ/+ and COM Short
[3]
e
Ohms
C14 C15 C15 C15 VΩ/+ and COM 40 Me
[3]
C15 C16 C16 C16 VΩ/+ and COM 6 V, 60 Hz
C16 C17 C17 C17 VΩ/+ and COM 60 V, 60 Hz
C17 C18 C18 C18 VΩ/+ and COM 600 V, 60 Hz
C18 C19 C19 C19 VΩ/+ and COM 6 V, 0 Hz
C19 C20 C20 C20 VΩ/+ and COM 60 V, 0 Hz
K
C20 C21 C21 C21 VΩ/+ and COM 600 V, 0 Hz
Set calibrator to standby, reconfigure leads, and program for amps output.
? N/A C22 N/A C22 A and COM 6 A, 60 Hz
[3]
A N/A C23 N/A C23 A and COM 6 A, 0 Hz
$
AC μamps
N/A N/A C22 N/A + and COM 600 μA, 60 Hz