Loading...
10, 11, 12 Multimeters
Service Manual
3-10
+
_
9V
BATTERY
+
CLIP TO BASE OF CONTACTS
ADJUST R4 FOR 3.997-4.003V DC
INPUT
RECEPTACLES
+
anao4f.eps
Figure 3-3. Calibration Adjustment