3 Fluke Corporation 1507/1503 Insulation Testers
Accurac
y (Model 1503)
Output Voltage Display Range Resolution Test Current
Accuracy
± (% of Rdg + Digits)
500 V
0.1 M to 20.00 M 0.01 M
(0 % to + 20 %)
20.0 M to 200.0 M 0.1 M 1 mA @ 500 k ± (2.0 % + 5)
200 M to 500 M 1 M
1000 V
0.1 M
to 200.0 M 0.1 M
1 mA @ 1 M ± (2.0 % + 5)
(0 % to + 20 %)
200 M to 2000 M 1 M
The following tables are a requirement for European labeling.
Measurement Intrinsic Uncertainty Operating Uncertainty
1
Volts ± (2.0 % + 3) 30 %
Earth Bond Resistance ± (1.5 % + 3) 30 %
Insulation Resistance Depends on test voltage and range. 30 %
See Insulation Test specifications.
1
This spec
ification comes from the standard and indicates the maximum amount allowable by the standard.
EN61557 influence variables and uncertainties
Earth Bond Resistance Uncertainty for Uncertainty for
Influence Variable Designation per EN61557 Insulation Resistance Earth Bond Resistance
Supply Voltage E2 5 % 5 %
Temperature E3 5 % 5 %
1
Specification confidence level 99 %.
EN61557 Specifications
The following tables can be used to determine the maximum or minimum display values considering maximum
instrument operating error per EN61557-1, 5.2.4.
Insulation resistance maximum and minimum display values
50 V 100 V 250 V 500 V 1000 V
Minimum Minimum Minimum Minimum Minimum
Limit V
alue
Display Limit Value Display Limit Value Display Limit Value Display Limit Value Display
V
alue
Value Value Value Value
0.05
0.0
7 0.05 0.07 0.05 0.07 0.05 0.07
0.06 0.08 0.06 0.08 0.06 0.08 0.06 0.08
0.07 0.09 0.07 0.09 0.07 0.09 0.07 0.09
0.0
8 0.10 0.08 0.10 0.08 0.10 0.08 0.10
0.0
9
0.
1
2
0.0
9
0.12 0.09 0.12 0.09 0.12
0.1 0.13 0.1 0.13 0.1 0.13 0.1 0.13 0.1 0.1
0.2 0.26 0.2 0.26 0.2 0.26 0.2 0.26 0.2 0.3
0.3
0.39
0.3
0.39 0.3 0.39 0.3 0.39 0.3 0.4
0.4 0.52 0.4 0.52 0.4 0.52 0.4 0.52 0.4 0.5
0.5
0.65
0.5
0.65
0.5 0.65 0.5 0.65 0.5 0.7
0.6 0.78 0.6 0.78 0.6 0.78 0.6 0.78 0.6 0.8
0.7 0.91 0.7 0.91 0.7 0.91 0.7 0.91 0.7 0.9
0.8 1.04 0.8 1.04 0.8 1.04 0.8 1.04 0.8 1.0
0.9
1
.
1
7
0.9
1.17 0.9 1.17 0.9 1.17 0.9 1.2
1.0 1.30 1.0 1.30 1.0 1.30 1.0 1.30 1.0 1.3
2.0 2.60 2.0 2.60 2.0 2.60 2.0 2.60 2.0 2.6
3.0
3.9
0 3.0 3.90 3.0 3.90 3.0 3.90 3.0 3.9
4.0 5.20 4.0 5.20 4.0 5.20 4.0 5.20 4.0 5.2
5.0
6.50
5.0
6.50 5.0 6.50 5.0 6.50 5.0 6.5
6.0 7.80 6.0 7.80 6.0 7.80 6.0 7.80 6.0 7.8

Other manuals for Fluke 1503

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1503 and is the answer not in the manual?

Fluke 1503 Specifications

General IconGeneral
Display type-
Housing colorGrey, Yellow
Maximum voltage600 V
Drop proof up to1 m
Operating altitude0 - 2000 m
Auto discharge time0.5 s
Operating vibration2 G
Overload protectionYes
Non-operating altitude0 - 12000 m
Maximum capacitive load1 µF
Storage temperature (T-T)-40 - 60 °C
Operating temperature (T-T)-20 - 55 °C
Insulation measurement range0.01 - 2000 MΩ
Insulation test voltage accuracy20 %
Operating relative humidity (H-H)0 - 95 %
Insulation measure accuracy (500V)1.5 %
International Protection (IP) codeIP40
Battery sizeAA
Number of batteries supported4
Weight and Dimensions IconWeight and Dimensions
Depth50 mm
Width100 mm
Height203 mm
Weight550 g