Electrical Specifications
31
1587 and 1587T Low-Pass Filter Voltage
Range Resolution
50 Hz to 60 Hz
±(% of Rdg + Digits)
60 Hz to 400 Hz
±(% of Rdg + Digits)
600.0 mV 0.1 mV
±(1 % + 3)
+(2 % + 3)
-(6 % - 3)
6.000 V 0.001 V
±(1 % + 3)
+(2 % + 3)
-(6 % - 3)
60.00 V 0.01 V
±(1 % + 3)
+(2 % + 3)
-(6 % - 3)
600.0 V 0.1 V
±(1 % + 3)
+(2 % + 3)
-(6 % - 3)
1000 V 1 V
±(2 % + 3)
+(2 % + 3)
-(6 % - 3)
1577 Accuracy
Range Resolution
50 Hz to 60 Hz
±(% of Rdg + Digits)
600.0 mV 0.1 mV
±(2 % + 3)
6.000 V 0.001 V
±(2 % + 3)
60.00 V 0.01 V
±(2 % + 3)
600.0 V 0.1 V
±(2 % + 3)
1000 V 1 V
±(2 % + 3)
AC Conversion ....................................................Inputs are ac-coupled and calibrated to the rms value of sine wave input. Conversions are
true-rms responding and specified from 5 % to 100 % of range. Input signal crest factor
can be up to 3 at up to 500 V, decreasing linearly to crest factor <= 1.5 at 1000 V. For
non-sinusoidal waveforms add ±(2 % reading + 2 % FS) typical, for a crest factor up to 3.
Input Impedance..................................................10 MΩ (nominal), <100 pF, ac-coupled