1587 FC/1587/1577
Calibration Manual
8
DC Voltage Measurement
Range Resolution
Specification (1587)
[1]
±(% of Rdg + Counts)
Specification (1577)
[1]
±(% of Rdg + Counts)
6.000 V dc 0.001 V
±(0.09 % + 2) ±(0.2 % + 2)
60.00 V dc 0.01 V
±(0.09 % + 2) ±(0.2 % + 2)
600.0 V dc 0.1 V
±(0.09 % + 2) ±(0.2 % + 2)
1000 V dc 1 V
±(0.09 % + 2) ±(0.2 % + 2)
[1] Specification applies to both positive and negative polarities.
Input Impedance ................................................. 10 MΩ (nominal), <100 pF
Normal Mode Rejection Ratio ............................. >60 dB @ 50 Hz or 60 Hz
Common Mode Rejection Ratio .......................... >120 dB @ dc, 50 Hz or 60 Hz (1 kΩ unbalance)
DC Millivolts Measurement
Range Resolution
Specification (1587)
±(% of Rdg + Counts)
Specification (1577)
±(% of Rdg + Counts)
600.0 mV dc 0.1 mV
±(0.1 % + 1) ±(0.2 % + 1)
DC and AC Current Measurement
Range Resolution
Specification (1587)
±(% of Rdg + Counts)
Specification (1577)
±(% of Rdg + Counts)
Burden Voltage
(Typical)
AC
45 Hz to 1000 Hz
400 mA 0.1 mA
±(1.5 % + 2)
[1]
±(2 % + 2)
[1]
2 mV/mA
60 mA 0.01 mA
±(1.5 % + 2)
[1]
±(2 % + 2)
[1]
DC
400 mA 0.1 mA
±(0.2 % + 2) ±(1.0 % + 2)
2 mV/mA
60 mA 0.01 mA
±(0.2 % + 2) ±(1.0 % + 2)
[1] 1 kHz bandwidth.
Overload ............................................................. 600 mA for 2 minutes maximum
Fuse Protection for mA Input .............................. 0.44 mA, 1000 V, IR 10 kA
AC Conversion .................................................... Inputs are ac-coupled and calibrated to the rms value of sine wave
input. Conversions are true-rms responding and specified from 5 % to
100 % of range. Input signal crest factor can be up to 3 up to 300 mA,
decreasing linearly to crest factor ≤1.5 at 600 mA. For non-sinusoidal
waveforms add +(2 % reading + 2 % FS) typical, for a crest factor up
to 3.
Ohms Measurement
Range Resolution
Specification (1587)
[1]
+(% of Rdg + Counts)
Specification (1577)
[1]
+(% of Rdg + Counts)
600.0 Ω 0.1 Ω
±(0.9 % + 2) ±(1.2 % + 2)
6.000 kΩ 0.001 kΩ
60.00 kΩ 0.01 kΩ
600.0 kΩ 0.1 KΩ
6.000 MΩ 0.001 MΩ
50.0 MΩ
[2]
0.01 MΩ ±(1.5 % + 3) ±(2.0 % + 3)
[1] Specifications apply from 0 % to 100 % of range.
[2] Up to 80 % relative humidity.
Overload Protection ............................................ 1000 V rms or dc
Open Circuit Test Voltage ................................... <8.0 V dc
Short Circuit Current ........................................... <1.1 mA
Diode Test (all 1587 models)
Diode Test Indication .......................................... Display voltage drop: 0.6 V at 1.0 mA nominal test current:
Accuracy ............................................................. ±(2 % Rdg + 3 counts)