EasyManua.ls Logo

Fluke 1587 FC - Page 18

Fluke 1587 FC
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1587 FC/1587/1577
Calibration Manual
10
Insulation Measurement (1587/1587 FC)
Output Voltage Display Range Resolution Test Current
Specification
±(% of Rdg + Counts)
50 V
(0 % to +20 %)
0.01 to 6.00 MΩ 0.01 MΩ
1 mA @ 50 k ±(3 % + 5)
6.0 to 50.0 MΩ 0.1 MΩ
100 V
(0 % to +20 %)
0.01 to 6.00 MΩ 0.01 MΩ
1 mA @ 100 k ±(3 % + 5)
6.0 to 60.0 MΩ 0.1 MΩ
60 to 100 MΩ 1 MΩ
250 V
(0 % to +20 %)
0.1 to 60.0 MΩ 0.1 MΩ
1 mA @ 250 k ±(1.5 % + 5)
60 to 250 MΩ 1 MΩ
500 V
(0 % to +20 %)
0.1 to 60.0 MΩ 0.1 MΩ
1 mA @ 500 k ±(1.5 % + 5)
60 to 500 MΩ 1 MΩ
1000 V
(0 % to +20 %)
0.1 to 60.0 MΩ 0.1 MΩ
1 mA @ 1 M
±(1.5 % + 5)
60 to 600 MΩ 1 MΩ
0.6 to 2.0 GΩ 100 MΩ ±(10 % + 3)
Insulation Measurement (1577)
Output Voltage Display Range Resolution Test Current
Specification
±(% of Rdg + Counts)
500 V
(0 % to +20 %)
0.1 to 60.0 MΩ 0.1 MΩ
1 mA @ 500 kΩ ±(2.0 % + 5)
60 to 500 MΩ 1 MΩ
1000 V
(0 % to +20 %)
0.1 to 60.0 MΩ 0.1 MΩ
1 mA @ 1 MΩ ±(2.0 % + 5)
60 to 600 MΩ 1 MΩ
Insulation Measurement (1587T)
Output Voltage Display Range Resolution Test Current
Specification
±(% of Rdg + Counts)
50 V
(0 % to +20 %)
0.01 to 6.00 MΩ 0.01 MΩ
1 mA @ 50 kΩ ±(3 % + 5)
6.0 to 50.0 MΩ 0.1 MΩ
100 V
(0 % to +20 %)
0.01 to 6.00 MΩ 0.01 MΩ
1 mA @ 100 kΩ ±(3 % + 5)
6.0 to 60.0 MΩ 0.1 MΩ
60 to 100 MΩ 1 MΩ
Insulation Measurement (1587C FC)
Output Voltage Display Range Resolution Test Current
Specification
±(% of Rdg + Counts)
50 V
(-10 % to +10 %)
0.01 to 6.00 MΩ 0.01 MΩ
1 mA @ 50 kΩ ±(3 % + 5)
6.0 to 50.0 MΩ 0.1 MΩ
100 V
(-10 % to +10 %)
0.01 to 6.00 MΩ 0.01 MΩ 1 mA @ 100 kΩ ±(3 % + 5)
250 V
(-10 % to +10 %)
0.1 to 60.0 MΩ 0.1 MΩ
1 mA @ 250 k ±(1.5 % + 5)
60 to 250 MΩ 1 MΩ
500 V
(0 % to +20 %)
0.1 to 60.0 MΩ 0.1 MΩ
1 mA @ 500 k ±(1.5 % + 5)
60 to 500 MΩ 1 MΩ
1000 V
(0 % to +20 %)
0.1 to 60.0 MΩ 0.1 MΩ
1 mA @ 1 M
±(1.5 % + 5)
60 to 600 MΩ 1 MΩ
0.6 to 2.0 GΩ 100 MΩ ±(10 % + 3)

Other manuals for Fluke 1587 FC