1587 FC/1587/1577
Calibration Manual
6
Electrical Specifications
Maximum Voltage Applied to
any Terminal and Common..................................... 1000 V
Fuse Protection for mA input .................................. 0.44 A, 1000 V, IR 10 kA
Batteries .................................................................. Four AA batteries (NEDA 15A or IEC LR6)
Battery Life .............................................................. Meter use 1000 hours; Insulation test use: Meter can perform at least
1000 insulation tests with fresh alkaline batteries at room temperature.
These are standard tests of 1000 V into 1 MΩ with a duty cycle of 5
seconds on and 25 seconds off.
Size ......................................................................... 5.0 cm H x 10.0 cm W x 20.3 cm L (1.97 in H x 3.94 in W x 8.00 in L)
Weight ..................................................................... 550 g (1.2 lb)
Altitude
Operating ........................................................ 2000 m
Storage ........................................................... 12,000 m
Over-Range Capability............................................ 110 % of range except for capacitance which is 100 %
Frequency Overload Protection .............................. 10
7
V Hz
Storage Temperature .............................................. -40 °C to 60 °C (-40 °F to 140 °F)
Operating Temperature ........................................... -20 °C to 55 °C (-4 °F to 131 °F)
Temperature Coefficient ......................................... 0.05 x (specified accuracy) per °C for temperatures <18 °C or >28 °C
(<64 °F or >82 °F)
Relative Humidity .................................................... Noncondensing
0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F)
0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F)
0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F)
Enclosure Protection ............................................... IEC 60529: IP40 (non-operating)
Safety
IEC 61010-1 .................................................... Pollution Degree 2
IEC 61010-2-033 ............................................ CAT IV 600 V / CAT III 1000 V
Wireless Radio with Adapter
Frequency Range ........................................... 2402 MHz to 2480 MHz
Output Power .................................................. <10 mW
Radio Frequency Certification ........................ FCC: T68-FBLE, IC: 6627A-FBLE
Electromagnetic Compatibility
International .................................................... IEC 61326-1:Portable Electromagnetic Environment; IEC 61326-2-2
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses
conductively-coupled radio frequency energy that is necessary for the
internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than
domestic and those directly connected to a low-voltage power supply
network that supplies buildings used for domestic purposes. There
may be potential difficulties in ensuring electromagnetic compatibility in
other environments due to conducted and radiated disturbances.
Emissions that exceed the levels required by CISPR 11 can occur
when the equipment is connected to a test object. The equipment may
not meet the immunity requirements of this standard when test leads
and/or test probes are connected.
Korea (KCC) ................................................... Class A Equipment (Industrial Broadcasting & Communication
Equipment)
Class A: Equipment meets requirements for industrial electromagnetic
wave equipment and the seller or user should take notice of it. This
equipment is intended for use in business environments and not to be
used in homes.
USA (FCC) ...................................................... 47 CFR 15 subpart B. This product is considered an exempt device per
clause 15.103.