Insulation Multimeters 
  Electrical Specifications 
7 
Electrical Specifications 
AC Voltage Measurement (all 1587 models) 
Range Resolution 
50 Hz to 60 Hz 
±(% of Rdg + Counts) 
60 Hz to 5000 Hz 
±(% of Rdg + Counts) 
600.0 mV  0.1 mV 
±(1 % + 3)  ±(2 % + 3) 
6.000 V  0.001 V 
±(1 % + 3)  ±(2 % + 3) 
60.00 V  0.01 V 
±(1 % + 3)  ±(2 % + 3) 
600.0 V  0.1 V 
±(1 % + 3)  ±(2 % + 3) 
[1]
 
1000 V  1 V 
±(2 % + 3)  ±(2 % + 3)
 
[1]
 
[1]  1 kHz bandwidth. 
AC Voltage Measurement with Low-Pass Filter (all 1587 models) 
Range Resolution 
50 Hz to 60 Hz 
±(% of Rdg + Counts) 
60 Hz to 400 Hz 
±(% of Rdg + Counts) 
600.0 mV  0.1 mV 
±(1 % + 3) 
+(2 % + 3) 
-(6 % + 3) 
6.000 V  0.001 V 
±(1 % + 3) 
+(2 % + 3) 
-(6 % + 3) 
60.00 V  0.01 V 
±(1 % + 3) 
+(2 % + 3) 
-(6 % + 3) 
600.0 V  0.1 V 
±(1 % + 3) 
+(2 % + 3) 
-(6 % + 3) 
1000 V  1 V 
±(2 % + 3) 
+(2 % + 3) 
-(6 % + 3) 
AC Voltage Measurement (1577) 
Range Resolution 
50 Hz to 60 Hz 
±(% of Rdg + Counts) 
600.0 mV  0.1 mV 
±(2 % + 3) 
6.000 V  0.001 V 
±(2 % + 3) 
60.00 V  0.01 V 
±(2 % + 3) 
600.0 V  0.1 V 
±(2 % + 3) 
1000 V  1 V 
±(2 % + 3) 
AC Conversion.................................................... Inputs are ac-coupled and calibrated to the rms value of sine wave 
input. Conversions are true-rms responding and specified from 5 % to 
100 % of range. Input signal crest factor can be up to 3 at up to 500 V, 
decreasing linearly to crest factor ≤1.5 at 1000 V. For non-sinusoidal 
waveforms add ±(2 % reading + 2 % FS) typical, for a crest factor up 
to 3. 
Input Impedance ................................................. 10 MΩ (nominal), <100 pF, ac-coupled 
Common Mode Rejection Ratio  
(1 kΩ unbalanced) .............................................. >60 dB at dc, 50 or 60 Hz