1587 FC/1587/1577 
Calibration Manual 
22 
Performance Test 
The following series of tests comprise a performance test for verifying the 
accuracy of the Meter (UUT) and its performance level. The performance test is 
recommended as an acceptance test for incoming inspection and as a calibration 
procedure for periodically ensuring the accuracy of the Meter. Fluke recommends 
running the performance test at least once a year. 
No adjustments are required during the performance test, and it is not necessary 
to open the case. If the Meter does not pass all parts of the performance tests, 
repair and/or calibration adjustment are required. A calibration adjustment 
procedure is given later in this manual. If significant repairs are required, contact 
Fluke as described toward the front of this manual. If user repairs are 
appropriate, refer to the list of user-replaceable parts toward the end of this 
manual 
Testing DMM Functions 
Use the following procedure to verify the accuracy of the measurement functions. 
1.  Connect the Calibrator to the J and COM input terminals on the UUT. 
2.  Turn the rotary switch to the function listed in each step of Table 3. 
3.  Apply the input level for each step listed in Table 3. 
4.  Compare the reading on the UUT display with the Display Reading in 
Table 3. 
If the display reading falls outside of the range shown in Table 3, the UUT 
does not meet specification. 
Table 3. DMM Accuracy Tests 
Step 
UUT 
Function 
UUT 
Range 
UUT 
Input Level 
UUT Display Reading 
1577 1587 FC/1587 
1  mV ac  600.0 mV  30 mV, 60Hz  29.4 to 30.6  29.4 to 30.6 
2  V ac  6.000V  3V, 60Hz  2.937 to 3.063  2.967 to 3.033 
3  V ac  6.000V  3V, 400Hz  n/a  2.937 to 3.063 
Push G (blue) for 1 second for low-pass filter band-pass test 
4  V ac  6.000V  3V, 400Hz  n/a  2.817 to 3.063 
5  V ac  6.000V  3V, 800Hz  n/a  1.680 to 2.670 
Push G (blue) for 1 second to remove low-pass filter 
6  V ac  60.00V  60V, 50Hz  58.77 to 61.23  59.37 to 60.63 
7  V ac  600.0V  600V, 50Hz  587.7 to 612.3  593.7 to 606.3 
8  V ac  1000V  1000V, 50 Hz  977 to 1023  n/a 
9  V ac  1000V  1000V, 1 kHz,   n/a  977 to 1023 
10  Hz, V ac  99.99 Hz  *10 Hz, 0.15V  n/a  9.98 to 10.02 
11  Hz, V ac  99.99 kHz  *95kHz, 0.15V  n/a  94.89 to 95.11