175, 177, & 179
Calibration Information
14
Table 3. Performance Tests 177/179
Meter Response
Test 5500 Output Lower Limit Upper Limit
300 mV 45 Hz 296.7 mV ac 303.3 mV ac
5 V 500 Hz 4.947 V ac 5.053 V ac
5 V 1 kHz 4.897 V ac 5.103 V ac
50 V 45 Hz 49.47 V ac 50.53 V ac
50 V 1 kHz 48.97 V ac 51.03 V ac
300 V 45 Hz 296.7 V ac 303.3 V ac
500 V 500 Hz 494.7 V ac 505.3 V ac
500 V 1 kHz 489.7 V ac 510.3 V ac
K
AC Volts
1000 V 45 Hz 987 V ac 1013 V ac
1 V 45 Hz 44.94 Hz 45.06 Hz
K
Hz
AC Volts
Frequency
1
5 V 50 kHz 49.94 kHz 50.06 kHz
5 V 4.993 V dc 5.007 V dc
300 V 299.5 V dc 300.5 V dc
1000 V 997 V dc 1003 V dc
L
DC Volts
1000 V -1003 V dc -997 V dc
3 V 45 kHz 44.94 Hz 45.06 Hz L
Hz
DC Volts
Frequency
1
30 V 50 kHz 49.94 kHz 50.06 kHz
30 mV 0 Hz 29.8 mV dc 30.2 mV dc
-300 mV -300.5 mV dc -299.5 mV dc
E
DC Millivolts
600 mV 599.3 mV dc 600.7 mV dc
0 °C -1.0 °C 1.0 °C
-40 °C -41.4 °C -38.6 °C
T
Temperature
1,3,4
(179 only)
400 °C 395.0 °C 405.0 °C
19 Ω 18.6 Ω 19.4 Ω e
Ohms
2
19 MΩ 18.68 M Ω 19.32 M Ω
N
Capacitance
1
0.90 μF 887 nF 913 nF
25 Ω Beeper On R
Continuity
250 Ω Beeper Off
O
Diode
1
2.0 V 1.978 V dc 2.022 V dc
1. Press the YELLOW button to select this function.
2. Does not include test lead resistance.
3. Be sure to use correct cable.
4. To ensure accuracy, the Meter and thermocouple adapter must be at the same temperature. Allow 20 minutes before performing the
temperature procedures.