EasyManua.ls Logo

Fluke 175 - Introduction; How to Contact Fluke; Safety Information

Fluke 175
22 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
True RMS Multimeters
Specifications
5
Accuracy ± ( [ % of Reading ] + [ Counts ] )
Function Range
1
Resolution Model 175 Model 177 Model 179
AC Volts
2,3
600.0 mV
6.000 V
60.00 V
600.0 V
1000 V
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
1.0 % + 3
(45 Hz to 500 Hz)
2.0 % + 3
(500 Hz to 1 kHz)
1.0 % + 3
(45 Hz to 500 Hz)
2.0 % + 3
(500 Hz to 1 kHz)
1.0 % + 3
(45 Hz to 500 Hz)
2.0 % + 3
(500 Hz to 1 kHz)
DC mV 600.0 mV 0.1 mV 0.15 % + 2 0.09 % + 2 0.09 % + 2
DC Volts 6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
0.15 % + 2
0.09 % + 2
0.09 % + 2
1000 V 1 V 0.15 % + 2 0.15 % + 2 0.15 % + 2
Continuity 600 Ω 1 Ω Meter beeps at < 25 Ω, beeper turns off at > 250 Ω; detects opens or shorts
of 250 μs or longer.
Ohms
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
50.00 MΩ
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
Diode test 2.400 V 0.001 V 1 % + 2
Capacitance
1000 nF
10.00 μF
100.0 μF
9999 μF
4
1 nF
0.01 μF
0.1 μF
1 μF
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
AC Amps
5
(True RMS)
(45 Hz to 1 kHz)
60.00 mA
400.0 mA
6.000 A
10.00 A
0.01 mA
0.1 mA
0.001 A
0.01 A
1.5 % + 3
1.5 % + 3
1.5 % + 3
1. All AC voltage and AC current ranges are specified from 5 % of range to 100 % of range.
2. Crest factor of 3 at full scale up to 500 V, decreasing linearly to crest factor 1.5 at 1000 V.
3. For non-sinusoidal waveforms, add -(2% reading + 2% full scale) typical, for crest factors up to 3.
4. In the 9999 μF range for measurements to 1000 μF, the measurement accuracy is 1.2 % + 2 for all models.
5. Amps input burden voltage (typical): 400 mA input 2 mV/mA, 10 A input 37 mV/A.

Other manuals for Fluke 175