Model 187 & 189
Getting Started
14
Setup Selections
The meter allows you to change the default operating
configuration of the meter by changing setup options
made at the factory. Many of these setup options affect
general meter operations and are active in all functions.
Others are limited to one function or group of functions.
To enter the Setup mode, turn the meter on and press
O T (SETUP). In the Setup mode, each press of
O T (SETUP) saves changes to the last selection
and steps to the next option.
Each setup option appears in the primary display in the
sequence shown in Tables 4 and 5.
The options in Table 4 are available only when the
preconditions are met. The options in Table 5 are
available for all functions. (When measuring dc volts,
none of the preconditions in Table 4 are required, and
only the selections shown in Table 5 will appear.)
To exit the Setup mode, Press O N (CANCEL).
Be sure to save your last selection by pressing O T
first.
Table 4. Function Specific Setup Selections
Selection Precondition Option Choices (
<
>
) Factory Default
000.0 °C
or
000.0 °F
Temperature (
w )
selected.
Temperature
offset adjust
000.0 ° to ±999.9 ° - Use q to increment or
decrement digit.
Use
<
>
to select digit. Selected digit
flashes.
000.0 °C (or °F)
L Int Model 189 only. Log interval
MM:SS - Use
q to increment or decrement
minute or second values.
Use
<
>
to select minute or seconds.
Selected values flash.
15:00
dbrEF AC volts (\ or
,) selected.
dB type dBm or dBV (m or V flashing) - Use
<
>
to
select.
dBV
dbrEF AC volts (\ or
,) and dBm
selected.
dBm reference 0001 to 1999 - Use q to increment or
decrement digit.
Use
<
>
to select digit.
0600