Service Manual
30
Table 8. Calibration Adjustment Steps for Firmware Version 1.10 (cont.)
Step
Rotary Switch
Position
Source Value
Settling Time
(seconds)
Calibrating Time
[1]
(seconds)
43 mA DC 400.0 mA, 0 Hz 2 1
44 mA AC 400.0 mA, 55 Hz 2 1
45 ADC 10.0 A, 0 Hz 2 1
46 AAC 10.0 A, 55 Hz 2 2
Connect the 5520A in a 4-Wire configuration and select COMP 2-Wire
To reduce settling time, apply 10 ohms to Meter before setting Rotary Switch to LoΩ
47 LoΩ (289) 10.000 Ω 2 2
48 LoΩ (289) 50.000 Ω 2 2
Disconnect 5520A 4-Wire configuration and turn COMP off
49 LoZ (289) 10.0 V, 0 Hz 2 1
50 mV AC (289) 500.00 mV, 55 Hz 2 5
[1] Meter is “Calibrating” after “Next” is pressed.